Chinese Journal of Diagnostics(Electronic Edition) ›› 2016, Vol. 04 ›› Issue (01): 48-51. doi: 10.3877/cma.j.issn.2095-655X.2016.01.014

Special Issue:

• Clinical Study • Previous Articles     Next Articles

The influence of room temperature and humidity on the measurement results of hepatitis C virus antibody using high-throughput ELISA method

Haixin Dong1, Chengqiang Jin1,(), Yuanming Zhao1, Jian Zhang1, Jing Sun1, Weixing Yu2   

  1. 1. Department of Clinical Laboratory, Affiliated Hospital of Jining Medical University, Jining 272029, China
    2. Department of Biotechnology, Jining Medical University, Jining 272067, China
  • Received:2015-10-05 Online:2016-02-26 Published:2016-02-26
  • Contact: Chengqiang Jin
  • About author:
    Corresponding author: Jin Chengqiang, Email:

Abstract:

Objective

To explore the influence of room temperature and humidity on the measurement results of hepatitis C virus antibody using high-throughput ELISA method.

Methods

To detect the S/CO values of HCV antibody standard using addcare automatic enzyme immunoassay instrument (ELISA 1100) under 24℃ and different humidity (30%RH, 40%RH, 50%RH, 60%RH, 70%RH, 80%RH, 90%RH) and 50%RH and different temperatures (16℃, 18℃, 20℃, 22℃, 24℃, 26℃, 28℃, 30℃, 32℃, 34℃, 36℃, 38℃, 40℃) respectively.

Results

The S/CO values of HCV-Ab standard showed significant difference under different temperature (16℃, 18℃, 20℃, 22℃, 24℃, 26℃, 28℃, 30℃, 32℃, 34℃, 36℃), and the S/CO value of standard was positively correlated with temperature.The S/CO value altered little between 36℃ and 38℃, but there was of significant difference between 38℃and 40℃.While the S/CO value of the standard gradually decreased with the rising of humidity(30%RH, 40%RH, 50%RH, 60%RH, 70%RH, 80%RH, 90%RH), which presented significant difference between different humidity (P<0.05), and S/CO value of the standard had a negative correlation with the humidity.

Conclusions

Room temperature and humidity have important effects on high-throughput ELISA detection results.High-throughput ELISA should been carried out in the environment with no more than 36℃ and low humidity.Comprehensive consideration is recommended for the automatic enzyme immunoassay instrument with operating temperature range from 18 to 25 degrees, humidity control between 35%RH~50%RH.

Key words: Temperature, Humidity, Enzyme-linked immunosorbent assay, Hepatitis C antibodies